The CG407 is a
single-channel oscilloscope ideal for
monitoring signals in your system.
| Parameter |
Value |
| Real-time sampling rate |
1MSPS (max) |
| Sensitivity |
1V/div - 5V/div in "1-2-5"
steps |
| Timebase |
10uS/div - 500uS/div in
"1-2-5" steps |
| Capture buffer size |
1024 bytes |
| Trigger modes |
Automatic/Normal/Single |
| Display size |
2.4" (61mm) |
| Display resolution |
320 pixels x 240 pixels |
| Input impedance |
1M ohm/20pF |
| ADC resolution |
12-bits |
The DataLogger section at the top of
the screen displays:-
- Vpp - Shows the difference between
the extreme Maximum and Minimum
values
- Vavr - The Average
reading is the mean vertical level of the
entire captured waveform
- Freq - The Frequency is
the inverse of the Period and is measured in
Hertz
Pressing
the [SELECT] button steps the user through the
various control settings displayed around the
edge of the screen, the currently active
setting being highlighted within a blue frame.
At each setting, pressing the [-] and [+]
buttons will step through the available
options.
Pressing
the [HOLD/OK] button will freeze the current
display allowing for closer inspection of the
input signal.
Pressing
the [HOLD/OK] button while in the lower left
corner of the screen, the user can toggle the
[AUTO SIZE] ([A.S]) function on and off. When
enabled, [A.S] will automatically resize the
input signal to fill the scope display area.
This is useful when you have a signal on the
display which doesn't optimally fit in any of
the [VOLTS/DIV] settings. The DataLogger still
displays the correct data for the input
signal.
.